3.2 Design
FIDAMAT 6
22 Operating Instructions, 01/2019, A5E00222135-04
Analyzer part
The analyzer part of the FIDAMAT 6-E consists of an oven in which the following
components are installed:
Flame ionization detector (FID)
Various restrictors
Heated sample gas filter (FIDAMAT 6-E only)
Outside the oven are:
2 pressure regulators
2 pressure sensors
Solenoid valves
Bulkhead unions 6 mm or 1/4"
Heated pump (FIDAMAT 6-E only)
Materials used
Table 3- 1 Materials used
Device component
Material
Gas path
Piping
Gas inlet
Gaskets
Sample gas restrictor
Auxiliary gas restrictors
Pump diaphragm (FIDAMAT 6-E only)
Pump head (FIDAMAT 6-E only)
Stainless steel 1.4571 / 1.4404
Stainless steel 1.4571 / 1.4404
Graphite
Stainless steel 1.4404
Stainless steel 1.4571 / 1.4404
PTFE
Stainless steel 1.4571 / 1.4404
Flame ionization detector (FID)
Nozzle
FID housing
Quartz
Stainless steel 1.4571 / 1.4404
3.2.2 Design of the enclosure
19" rack unit with 4 HU
For installation in a hinged frame
In cabinets, with telescopic or support rails
Front panel can be swiveled downward for service purposes (for example, for connecting
a notebook via RS 485)
Gas connections for sample gas inlet (sample) and gas outlet (exhaust), as well as
combustion gas and combustion air, pipe union for pipes with outer diameter of 6 mm or
1/4"