Functions
8.3 Calibration
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
83
8.3.5 Total / single calibration (function 23)
Figure 8-7 Select calibration method
With this function, select between the total and single calibration of the measuring ranges.
Total calibration means that you adjust one "leading measuring range" and all other
measuring ranges are calibrated using the switching ratio.
If the total calibration, as shown on the function screen, is not activated, you must calibrate
every range individually.
8.3.6 AUTOCAL (function 24)
Note
You can only make use of automatic calibration (AUTOCAL) if your device contains
additional electronics (option board). If it does not, a corresponding message will appear on
the display when the AUTOCAL function is called.
Automatic calibration can only be started when the device is in "Measuring" operating state!
Figure 8-8 AUTOCAL function