EasyManuals Logo

Siemens FIDAMAT 6 User Manual

Siemens FIDAMAT 6
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
Commissioning
6.3 Commissioning and operation
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
59
4. Combustion air test
The device checks the combustion air pressure.
5. Igniting
The flame is automatically ignited. The flame burns when the flame temperature is
approximately 20 °C higher than the oven temperature (can be called under function 2
"Diagnostic values").
During the warm-up phase, "WARMUP" appears in the bottom line of the measured value
display, alternately with the current warm-up step.
FIDAMAT 6-E: When the device has detected that the flame is burning, the pump is
switched on and the analyzer is set to "Measuring" operating state.
FIDAMAT 6-G: When the device has detected that the flame is burning, the control
system sets the device to "Standby" operating state. The control system switches to the
"Measuring" operating state only when you turn on the zero gas, calibration gas or
sample gas using function 62 (or function 72 "Binary inputs") and the sample gas
pressure (consisting of "zero gas, calibration gas or sample gas" in conjunction with
"combustion air") has exceeded the value 485 hPa (can be set using function 92).

Table of Contents

Other manuals for Siemens FIDAMAT 6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens FIDAMAT 6 and is the answer not in the manual?

Siemens FIDAMAT 6 Specifications

General IconGeneral
BrandSiemens
ModelFIDAMAT 6
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals