Operation
7.1 User interface and operating modes
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
67
Device operating modes
Table 7- 2 Operating modes of the FIDAMAT 6
Mode
Properties
Remarks
Coded display mode
Measured value display is
shown
Protected submenus can
only be reached by entering
a code
The current operating state
of the device (except for
"Measuring") is displayed in
the bottom line
Function control not active
The device only supplies relia-
ble measured values in this
mode, provided it is in "Measur-
ing" operating state. You reach
this mode from operator control
mode by pressing the [MEAS]
key twice.
Decoded display mode
Measured value display is
shown
The submenus protected by
the entered code are acces-
sible
The current operating state
of the device (except for
"Measuring") is displayed in
the bottom line
Function control active
Measured value can be
influenced
You reach this mode from oper-
ator control mode by pressing
the [MEAS] key once and con-
firming or discarding the entries
that were made.
Operator control mode
Menu or function is dis-
played
Settings and inputs can be
made
Function control active
Measured value can be
influenced
You reach this mode from
"Coded display mode" by enter-
ing the code of the correspond-
ing operation level.
You configure/calibrate the
device in this mode.