8.4 Measuring ranges
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
91
Measuring range types
This yields the permissible measuring range types shown in the figure below:
Figure 8-18 Measuring range types
Two measuring range types are distinguished:
Type A:
The measuring range end value must be smaller than the measuring range end value
following it. The top measuring range limit thus becomes higher with every measuring range.
mba
Start of measuring range
mbe
End of measuring range
UU
Lower switchover point: Select smaller measuring range
OU
Upper switchover point: Select larger measuring range
Figure 8-19 Measuring range type A
The following applies to the measuring range switchover: