Interrupts, error and system alarms
12.2 Errors
SITRANS P DS III with PROFIBUS PA
Operating Instructions, 06/2013, A5E00053276-07
189
Table 12- 2 Error message
Display
PDM display
Cause
Measure
F_001
-
Local operation blocked.
Remove write protection.
F_003 - Changes to the bus address and
device operating type are not
possible, since the device is in
communication with a class 1
master.
End communication with class
1 master.
F_004 - Display overflow. Check settings of physical unit
and position of decimal point,
and adjust to the current
measured value.
F_005
-
Value is read-only.
-
F_006 - Correction not successful. Check calibration span and
repeat procedure.
F_007 - After zero-point calibration,
measurements no longer
possible in entire measurement
range.
Check measurement range,
decrease correction if
necessary.
F_008 - Local operation blocked by
SIMATIC PDM.
Use SIMATIC PDM to set the
"Local operation" parameter to
"released".
See also
Status (Page 162)