Operator control functions via PROFIBUS
7.4 Settings
SITRANS P DS III with PROFIBUS PA
Operating Instructions, 06/2013, A5E00053276-07
111
● Create an assignment between the measured mass value and the output value by setting
these parameters:
Function block 1 - analog input
Measured value (primary variable)
Start of scale value:
Full scale value:
king range"
Start of scale value:
Full scale value:
Unit of mass [kg, g, t ...]
as "measured value scaling" * density
as "measured value scaling" * density
You can also adjust the output to another process value. Assign the required unit to the
parameters under the rubric output scaling. Assign the start of scale and full scale values
according to the measured value scaling.
Adjusting to a desired process value (Page 114)
● Select the required configuration with the configuration tool:
Table 7- 6 Desired configuration for flow measurement
Totalizer output Volume or mass that has flowed over a time
Output, totalizer output Current flow/time,
volume or mass that has flowed over a time
segment,
reset totalizer output (with SIMATIC PDM)
Output, totalizer output, reset totalizer output Current flow/time,
volume or mass that has flowed over a time
segment,