Technical data
13.4 Measuring accuracy
SITRANS P DS III with PROFIBUS PA
200 Operating Instructions, 06/2013, A5E00053276-07
Measuring accuracy (as per EN 60770-1) gauge pressure
PROFIBUS PA and FOUNDATION
Fieldbus
Reference conditions
• Rising characteristic curve
• Start of scale value 0 bar
• Seal diaphragm stainless steel
• Measuring cell with silicone oil filling
• Room temperature 25 °C (77 °F)
• Measuring span ratio r
r = max. measuring span/set
–
Measurement deviation with limit
setting, including hysteresis and
Linear characteristic curve
• r ≤ 10
≤ (0.0029 • r + 0.071) % –
• 10 < r ≤ 30
≤ (0.0045 • r + 0.071) % –
• 30 < r ≤ 100
≤ (0.005 • r + 0.05) % –
Included in the measurement deviation
Included in the measurement deviation
Settling time T
63
without electrical
Approx. 0.2 s
Long-term drift at ±30 °C (±54 °F) In 5 years
• 1- to 4-bar measuring cell
In 5 years
• 16 to 700 bar measuring cell
In 5 years
Effect of ambient temperature
• at -10 … +60 °C (14 … 140 °F)
≤ (0.08 • r + 0.1) % ≤ 0.3%
• at -40 … -10°C and
+60 … +85°C
(-40 … 14°F and 140 … 185°F)
≤ (0.1 • r + 0.15) % per 10 K ≤ 0.25 % per 10 K
Effect of mounting position ≤ 0.05 mbar g (0.000725 psi g) per 10° inclination
correction via zero offset
Effect of auxiliary power supply In percent per change in voltage
–
Measuring value resolution
-5
of the rated measuring range