3.6 Principle of operation
SITRANS P DS III with PROFIBUS PA
Operating Instructions, 06/2013, A5E00053276-07
27
3.6.2
Operation of the electronics
Description
Measuring cell sensor
Buttons (local operation)
Measuring amplifier
Display
Analog-to-digital converter
Power supply
Microcontroller
DP/PA coupler or DP/PA link
Electrical isolation
Bus master
Each with an EEPROM in the measuring cell
and in the electronics
p
e
Input variable
PROFIBUS PA interface
Figure 3-6 Principle of operation of the electronics with PA communication
Function
The inlet pressure is converted into an electrical signal by the sensor
.
This signal is amplified by the measuring amplifier
and digitized in an analog-to-digital
converter
.
The digital signal is analyzed in a microcontroller
and corrected with regard to linearity
and thermal characteristics.
The signal is available at an electrically isolated PA interface
on the PROFIBUS PA.