Technical data
13.2 Input point
SITRANS P DS III with PROFIBUS PA
Operating Instructions, 06/2013, A5E00053276-07
193
gauge pressure input, with flush mounted diaphragm
PROFIBUS PA and FOUNDATION Fieldbus
Span (continuously
adjustable) or measuring
range, max. operating
pressure and max. test
pressure
Span Maximum
operating
pressure
Maximum test
pressure
Measuring
range
Maximum
operating
pressure
Maximum test
pressure
0.01 …
1 bar g (0.15
4 bar g
(58 psi g)
6 bar g
(87 psi g)
1 bar g
(14.5 psi g)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
0.04 …
4 bar g (0.58
7 bar g
(102 psi g)
10 bar g
(145 psi g)
4 bar g
(58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
0.16 …
16 bar g (2.3
21 bar g
(305 psi g)
32 bar g
(464 psi g)
16 bar g
(232 psi g)
21 bar g
(305 psi g)
32 bar g
(464 psi g)
0.6 …
63 bar g (9.1
67 bar g
(972 psi g)
100 bar g
(1450 psi g)
63 bar g
(914 psi g)
67 bar g
(972 psi g)
100 bar g
(1450 psi g)
• Measuring cell with
silicone oil filling
100 mbar a (1.45 psi a)
• Measuring cell with
inert liquid
100 mbar a (1.45 psi a)
• Measuring cell with
neobee
100 mbar a (13.05 psi a)
100% of max. measuring range
Absolute pressure input, with flush-mounted diaphragm
PROFIBUS PA and FOUNDATION Fieldbus
Measured variable Absolute pressure
Span (continuously
adjustable) or measuring
range, max. operating
pressure and max. test
pressure
Span Maximum
operating
pressure
Maximum test
pressure
Measuring
range
Maximum
operating
pressure
Maximum test
pressure
43 …
1300 mbar a
(17 ...
2.6 bar a
(37.7 psi a)
10 bar a
(145 psi a)
1.3 bar a
(18.9 psi a)
2.6 bar a
(37.7 psi a)
10 bar a
(145 psi a)
160 …
5000 mbar a
(2.32 …
10 bar a
(145 psi a)
30 bar a
(435 psi a)
5 bar a
(72.5 psi a)
10 bar a
(145 psi a)
30 bar a
(435 psi a)