7-12
764333-098
2. Press OK/RESET button, display shows:
Aout1
signal = __
3. Press RIGHT arrow once to move cursor to under-
score position to enter signal value from conductiv-
ity sensor corresponding to analog output 1. Dis-
play shows:
Aout1 signal = _
_
4. Press UP arrow until A appears on display:
Aout1 signal = A
5. Press OK/RESET button, display shows:
Aout1 Min = 0
.000_
6. Use ARROWS to enter 100.0µ. Display shows:
Aout1 Min = 100.0µ
7. Press OK/RESET button, display shows:
Aout1 Max = 0.000_
8. Use ARROWS to enter 40.00µ, display shows:
Aout1 Max= 40.000µ.
9. Press OK/RESET button, display shows:
Aout1 Mid = 0.000_
10.Press OK/RESET button, display shows:
1 : If err set M
ax
11.Press UP arrow until Min appears on display:
1 : If err set M
in
12.Press OK/RESET button, display shows:
Save Changes Yes
13.Press OK/RESET button to confirm modifications.
• Channel 2 Adjustment:
1. Press OUTPUT button. Display shows:
Output: A
nalog
2. Press OK/RESET button, display shows:
Aout1
signal = __
3. Press UP arrow until 2 appears on display:
Aout2
signal = __
4. Press RIGHT arrow once to move cursor to under-
score position to enter signal value from pH sensor
corresponding to analog output 2. Display shows:
Aout2 signal = _
_
5. Press UP arrow until B appears on display:
Aout2 signal = B
6. Press OK/RESET button, display shows:
Aout2 Min = 0.000_
7. Use ARROWS to enter 1.00, display must show:
Aout2 Max = 1.00_
8. Press OK/RESET button, display shows:
Aout2 Max = 0.000_
9. Use ARROWS to enter 15.00, display must show:
Aout2 Max = 15.00_
10.Press OK/RESET button, display shows:
Aout2 Mid = 0.000_
11.Press OK/RESET button, display shows:
2 : If err set M
ax
12.Press UP arrow until Min appears on display:
2 : If err set M
in
13.Press OK/RESET button, display shows:
Save Changes Yes
14.Press OK/RESET to confirm modifications.
7.13.5 Thornton Display Adjustment:
1. Display must show:
A*.**S B*.**pH
2. Press UP arrow until right information appears on
display.
7.14 HOW TO REPLACE pH SENSOR (pH
NEUTRALIZING SYSTEM OPTION)
NOTE: pH Controller must be calibrated after sensor
replacement.
1. Lock unit electrical disconnect switch in OFF
position.
2. Locate and drain neutralizing pump piping.
3. Locate pH sensor, unplug connector and remove.
4. Remove cap on new pH probe, install and plug
connector.
5. Perform calibration (refer to Section 2.5, Thornton
Controller).
7.15 RTD REPLACEMENT
There are four RTDs:
•Sump RTD (Sump RTD) is located in chamber
sump;
• Alkaline tank and/or acid tank RTD is locted on
side of respective tank;