7-11
Troubleshooting Operator Manual 920509-645
16. Alarm sounds and screen dis-
plays:
EMERGENCY STOP PRESSED
and printer prints:
ALARM: HH:SS:MM
EMERGENCY STOP
PRESSED
EMERGENCY STOP button pressed.
1. Press ACK touch pad to acknowledge alarm.
2. Pull out EMERGENCY STOP buttons on both control sides.
3. Press RESUME CYCLE touch pad to resume cycle operation or
press ABORT CYCLE touch pad to abort cycle operation.
4. If situation reoccurs, call STERIS.*
17. Alarm sounds and screen dis-
plays:
COOLDOWN TANK TOO LONG
TO FILL
and printer prints:
ALARM: HH:SS:MM
COOLDOWN TANK TOO LONG
TO FILL
1. Press ACK touch pad to acknowledge alarm.
2. Self-cleaning screen clogged – remove debris from screen.
3. Pump suction strainer clogged – clean strainer.
4. Press RESUME CYCLE touch pad to resume cycle operation.
5. If situation reoccurs, call STERIS.*
18. Alarm sounds and screen dis-
plays:
COOLDOWN TANK TOO LONG
TO COOL
and printer prints:
ALARM: HH:SS:MM
COOLDOWN TANK TOO LONG
TO COOL
1. Press ACK touch pad to acknowledge alarm.
2. Cold water supply valve not fully open – open building and washer
supply valves.
3. Press RESUME CYCLE touch pad to resume cycle operation.
4. If situation reoccurs, call STERIS.*
19. Alarm sounds and screen dis-
plays:
COOLDOWN TANK TOO LONG
TO DRAIN
and printer prints:
ALARM: HH:SS:MM
COLLDOWN TANK TOO LONG
TO DRAIN
1. Press ACK touch pad to acknowledge alarm.
2. Drain valve not fully open – open valve.
3. Press RESUME CYCLE touch pad to resume cycle operation.
4. If situation reoccurs, call STERIS.*
* Service Charges may be incurred. Consult your warranty for details.
Table 7-2. Fault Chart (Cont’d)
PROBLEM POSSIBLE CAUSE AND CORRECTION