3. Automated wash
• Program the washer using the following parameters:
Phase Recirculation
Time
Water
Temperature
Detergent Type and
Concentration
(if applicable)
Pre Wash 2 minutes Cold tap
water
N/A
Enzyme
Wash
2 minutes Hot tap water Enzymatic Detergent
Wash 1 2 minutes Set point
(66 ˚C)
Non-enzymatic
Detergent
4
Rinse 1 2 minutes Hot tap water N/A
Dry Phase 7 minutes 115 ˚C N/A
• If necessary, use pressurized air to aid in drying. Visually inspect each
device for cleanliness.
4
Prolystica® 2x Neutral Detergent is validated for cleaning ecacy.
Low Level Disinfection (optional)
1. Clean and prepare the camera head and cable as recommended in this
user guide. Ensure the soaking cap is installed.
2. Disinfect the device in a disinfecting solution that has one of the
following active Ingredients:
• ≥ 2.4% glutaraldehyde
5
with a minimum soaking time of 45 min at
25°C
• ≥ 3.4% glutaraldehyde
6
with a minimum soaking time of 20 min at
25°C
• ≥ 0.55% ortho-phthalaldehyde
7
with a minimum soaking time of
12min at 25°C.
3. Prepare the disinfecting solution according to the manufacturer’s
instructions.
4. Thoroughly rinse and ush all parts and lumens with running,
demineralized water to remove the disinfectant.
5. Dry all parts with a lint-free towel immediately after rinsing.
5
CIDEX Activated® is validated for disinfection ecacy.
6
CIDEX Plus® is validated for disinfection ecacy.
7
CIDEX® OPA is validated for disinfection ecacy.