Field or control Description
Display Sets the display of the mark icons on or off. If you have multiple searches defined, the control turns off
just the marks for the selected search.
Sets the display of the mark icons on or of
f.
Search Type Set to Timeout.
Source Lists the source channel or waveform to use to trigger or search. Types that require multiple inputs will
replace this control with a different source definition control.
Mark When
Stays High: The signal stays above the specified threshold level longer than the specified time.
Stays Low: The signal stays below the specified threshold level longer than the specified time.
Either: The signal stays above or below the specified threshold level longer than the specified
time.
Threshold Sets the amplitude level that the signal must pass through to be considered a valid transition.
Set to 50% Sets the threshold at 50% of the measured signal transition range. 50% is calculated as (Top +
Bottom)/2.
Time Limit Sets the time period condition to be met.
Copy Trigger Settings to
Search
Sets the search criteria to match the current oscilloscope trigger settings. If the trigger settings are not
valid in Search, this control is either not available or grayed out.
Copy Search Settings to
Trigger
Sets the current oscilloscope trigger settings to match the search criteria.
Other search types
Bus Search configuration menus on page 135
Edge Search configuration menu on page 152
Logic Search configuration menu on page 153
Pulse Width Search configuration menu on page 155
Rise/Fall T
ime Search configuration menu on page 156
Runt Search configuration menu on page 158
Setup and Hold Search configuration menu on page 159
Analog Channel configuration menu
Use the Analog Channel configuration menu to set up analog channel vertical settings, probe settings, deskew settings, external
attenuation, and alternate units for analog channel inputs.
To open an analog channel configuration menu, double-tap an analog channel badge. The following text describes analog channel
settings. For digital channel settings, see Digital channel configuration menu on page 173.
Vertical Settings panel, fields and controls
Field or control Description
Display
Toggles display of the channel On and Off.
Table continued…
Menus and dialog boxes
3 Series Mixed Domain Oscilloscope Printable Help 161

Table of Contents

Other manuals for Tektronix MDO34

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO34 and is the answer not in the manual?

Tektronix MDO34 Specifications

General IconGeneral
Analog Channels4
Sample Rate2.5 GS/s
Record Length10 Mpoints
Mixed Domain AnalysisYes
Waveform Capture Rate>280, 000 wfms/s
Spectrum Analyzer Frequency Range9 kHz to 3 GHz
Vertical Resolution8 bits
Input CouplingAC, DC, GND
Operating Temperature0°C to +50°C
Digital Channels16
ConnectivityUSB, LAN, HDMI
Arbitrary Function Generator50 MHz
Trigger TypesEdge, Pulse Width, Runt, Logic, Setup/Hold, Rise/Fall Time, Video
Input Impedance1 MΩ
Maximum Input Voltage300 Vrms
Dimensions140 mm
Weight4.5 kg
Bandwidth100 MHz, 200 MHz, 350 MHz, 500 MHz
Display10.1-inch capacitive touch HD display
Voltmeter4-digit