Customizing measurements
Customizing measurements
After adding a measurement, you can customize the measurement for more precise results by using gating or setting reference levels.
T
o customize measurements, double-tap a Measurement badge in the Results bar to open the Measurement configuration
menu on page 111.
See the following topics for more information.
Reference levels determine how time-related measurements are taken. See Set measurement reference levels on page 94.
Gating confines the measurement to a certain portion of a waveform. See Set measurement gates on page 94.
Set measurement reference levels
Use this procedure to set measurement reference levels.
About this task
Reference levels are set in the Global Measurement Settings panel of the Measurements configuration menu. See Measurement
configuration menu on page 111.
Before you begin
Prerequisite: To set measurement reference levels you must be taking a measurement. See Add a measurement on page 72.
Procedure
1. Double-tap a Measurement badge.
The Measurement configuration menu is displayed.
2. Tap the Global Measurement Settings panel.
3. Tap Reference Levels and select either % or Units.
% sets the High, Mid, and Low reference levels as percentages of the calculated High and Low signal levels. Tap the High Ref,
Mid Ref, or Low Ref fields and set the level using the assigned multipurpose knob. Or double-tap the field and use the virtual
keypad to enter a value.
Units sets the High, Mid, and Low reference levels to specific signal levels. Tap the High Ref, Mid Ref, or Low Ref fields and set
the level using the assigned multipurpose knob. Or double-tap the field and use the virtual keypad to enter a value.
4. Tap High-Low Method and select the method.
Auto automatically chooses the method.
Hysteresis is best for pulses.
Min-Max is best for all other waveforms.
5. Tap anywhere outside the Measurement configuration menu to close it.
Set measurement gates
Use this procedure to specify which portion of your waveform is used to take measurements.
About this task
Gating is set in the Global Measurement Settings panel of the Measurements configuration menu. See Measurement configuration
menu on page 111.
Customizing measurements
3 Series Mixed Domain Oscilloscope Printable Help 94

Table of Contents

Other manuals for Tektronix MDO34

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO34 and is the answer not in the manual?

Tektronix MDO34 Specifications

General IconGeneral
Analog Channels4
Sample Rate2.5 GS/s
Record Length10 Mpoints
Mixed Domain AnalysisYes
Waveform Capture Rate>280, 000 wfms/s
Spectrum Analyzer Frequency Range9 kHz to 3 GHz
Vertical Resolution8 bits
Input CouplingAC, DC, GND
Operating Temperature0°C to +50°C
Digital Channels16
ConnectivityUSB, LAN, HDMI
Arbitrary Function Generator50 MHz
Trigger TypesEdge, Pulse Width, Runt, Logic, Setup/Hold, Rise/Fall Time, Video
Input Impedance1 MΩ
Maximum Input Voltage300 Vrms
Dimensions140 mm
Weight4.5 kg
Bandwidth100 MHz, 200 MHz, 350 MHz, 500 MHz
Display10.1-inch capacitive touch HD display
Voltmeter4-digit

Related product manuals