3. Double-tap the Enter License Key field to open the virtual keyboard.
4. Use the virtual keyboard to enter the license key
. The key is not case sensitive.
5. Tap OK. The instrument installs the license and returns to the About screen.
6. Power cycle the instrument to enable the installed option(s).
7. Return to the Help > About menu to verify that the Enabled status for the option is Yes.
8. If you installed a bandwidth upgrade, remove the model/bandwidth label from the lower-left corner of the front panel and install the
new model/bandwidth label that your received as part of the upgrade purchase. If you installed a bandwidth upgrade, run signal path
compensation (Utility > Calibration > Run SPC).
Option upgrades
3 Series Mixed Domain Oscilloscope Printable Help 31

Table of Contents

Other manuals for Tektronix MDO34

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO34 and is the answer not in the manual?

Tektronix MDO34 Specifications

General IconGeneral
Analog Channels4
Sample Rate2.5 GS/s
Record Length10 Mpoints
Mixed Domain AnalysisYes
Waveform Capture Rate>280, 000 wfms/s
Spectrum Analyzer Frequency Range9 kHz to 3 GHz
Vertical Resolution8 bits
Input CouplingAC, DC, GND
Operating Temperature0°C to +50°C
Digital Channels16
ConnectivityUSB, LAN, HDMI
Arbitrary Function Generator50 MHz
Trigger TypesEdge, Pulse Width, Runt, Logic, Setup/Hold, Rise/Fall Time, Video
Input Impedance1 MΩ
Maximum Input Voltage300 Vrms
Dimensions140 mm
Weight4.5 kg
Bandwidth100 MHz, 200 MHz, 350 MHz, 500 MHz
Display10.1-inch capacitive touch HD display
Voltmeter4-digit