Measure Protein Pierce 660
Thermo Scientific NanoDrop One User Guide 125
Protein Pierce 660 standard curve screen
The standard curve screen shows graphically the relationship between the measured standards,
the calculated standard curve, and the measured absorbance and calculated concentration for
a selected sample. A horizontal line connects the sample absorbance value on the Y-axis to the
standard curve. A vertical line connects that point to the sample concentration value on the
X-axis.
The R
2
value indicates how well the standard curve fits the standard data points (1.0 is a
perfect fit; that is, all points lie exactly on the curve).
Tap row to
select sample
and update
graph; tap more
rows to overlay
abs/conc lines
for up to five
samples. Press
and hold
sample row for
measurement
details.
Drag tab
down/up to see
more/less
sample data
Pinch and zoom to adjust axes;
double-tap to reset
Standard curve
Menu of options;
tap to open
Protein concentration
for selected sample
Absorbance
at 660 nm
Page control; swipe screen left or
right to view next or previous screen
Curve type
R
2
value (1.0
equals perfect fit)
Red squares indicate
sample data points
White circles indicate
standard data points
Selected sample