Maintaining Your Instrument
Instrument Diagnostics
Thermo Scientific NanoDrop One User Guide 291
Instrument Diagnostics
Every 6 months, run the following performance and quality checks to verify instrument
operation.
Intensity Check
Run Intensity Check every 6 months to verify operation of the instrument’s internal
components. The test measures the intensity of light from the xenon source through the
instrument to verify that throughput, wavelength accuracy, and bias are within specifications.
If the instrument has a cuvette holder (NanoDrop One
C
model only), the test is automatically
repeated using the cuvette optical path.
Supplies needed
lint-free laboratory wipes
To run intensity check
1. Lift the instrument arm and clean the upper and lower pedestal with a new laboratory
wipe.
2. For the NanoDrop One
C
model instrument, remove any cuvette from the cuvette holder.
3. Lower the arm.
4. From the instrument home screen, tap (Diagnostics) and then tap
Intensity Check.
Intensity Check Performance Verification
Pedestal Image Check

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo NanoDrop One and is the answer not in the manual?

Thermo NanoDrop One Specifications

General IconGeneral
Wavelength Range190-850 nm
Pathlength1 mm and 0.2 mm (Auto-ranging)
Detection Limit2 ng/µL (dsDNA)
Light SourceXenon flash lamp
Maximum Concentration15, 000 ng/µL (dsDNA)
Detector2048-element linear silicon CCD array
Measurement Time< 5 seconds
Weight2.0 kg (4.4 lbs)
Nucleic Acid QuantitationYes
Protein QuantitationYes
Display7-inch capacitive touch screen
Sample Volume1-2 μL
Absorbance Precision±0.002 A
Operating Voltage100 to 240 V AC
Dimensions (W x D x H)20 cm x 25 cm x 15 cm (8 in. x 10 in. x 6 in.)
SoftwareThermo Scientific™ software
ConnectivityUSB, Ethernet, Wi-Fi
Operating Temperature15 - 35°C
Dimensions20 cm x 25 cm x 15 cm (8 in. x 10 in. x 6 in.)
Absorbance Range0 - 550 A (10 mm equivalent)