Measure OD600
Thermo Scientific NanoDrop One User Guide 135
Setting Available Options Description
Absorbance correction Absorbance value
between 0 and 300 A
User-defined absorbance correction. Enter absorbance correction
for displayed spectrum. This can be useful, for example, to
correct baseline offset caused by any difference between the
media solution used to blank the instrument and media used to
suspend the cell culture sample, and because scattered light
generally produces an offset.
Absorbance correction value is subtracted from absorbance values
at all wavelengths in sample spectrum. (All displayed absorbance
values are corrected values.)