Set Up the Instrument
Thermo Scientific NanoDrop One User Guide 191
Show connectivity status
– tap on instrument Home screen to open System Status box
–tap OK to exit System Status
Operating Specifications
The instrument operates reliably when the room environment meets these specifications:
• operating temperatures: 5 °C - 35 °C (41 °F - 95 °F)
• relative humidity (non-condensing): 20-80%
Locate the instrument away from air vents and exhaust fans to minimize evaporation.
After the instrument is installed, you can leave it turned on.
Related Topics
• Safety and Operating Precautions
• Instrument Models and Features
Location of database where instrument is currently
storing data (Local (instrument) or Connected PC)
Wi-Fi status
Bluetooth
status
Note If operating the instrument at the low end of the recommended humidity range, use
adequate sample volume to avoid evaporation.