4
Maintaining Your Instrument
Instrument Diagnostics
Thermo Scientific NanoDrop One User Guide 297
After the tenth measurement, a message indicates whether the instrument passed or failed
performance verification:
8. If the instrument failed, immediately repeat step 6 using ten 2 μL aliquots of the PV-1
solution.
9. When finished, tap End Experiment and clean the pedestals with 3–5 μL DI H
2
O.
After the test is completed, the results are available from the Data Viewer (see example
below). See Manage identifiers on the instrument for details.
To interpret performance verification results
If your instrument failed performance verification and you repeated ten measurements using
2uL aliquots, contact us.
Related Topics
• PV-1 Performance Verification Solution
• Intensity Check
• Pedestal Image Check