Safety Information
Thermo Scientific NanoDrop One User Guide 307
Power Cords
Be sure to use an appropriate grounded power cord for your electrical service. If the power
cord received is not appropriate for the electrical system in your location, or if the power cord
becomes damaged, contact us.
Power Line Conditioning Accessories
A UPS reduces the probability of a system shutdown if power is lost elsewhere in the building.
Power line conditioners (which ensure that your service is free from sags, surges or other line
disturbances) also are available in the U.S.A. from us for 120 volt operation. Line conditioners
for 220 volt operation can be purchased locally. Contact technical support for information
about power conditioners and UPS.
Electrical Service Specifications
The following table lists the specifications for electrical service. Contact our service
representative in your area if you have questions about the requirements.
Power Consumption
Generally, 50% more power should be available than the entire system (including accessories)
typically uses. Maximum power consumption and heat dissipation specifications for the
spectrometer and accessories are shown below. The values are approximate.
Requirements Specifications
Input current 5.0 A (max.)
Input voltage 100-240 VAC
Line frequency 50-60 Hz
Line disturbances Sags, surges or other line disturbances must not
exceed 10% of input voltage (even for a half cycle).
Noise < 2 V (common mode)
< 20 V (normal mode)
Item Power Consumption Max. Heat Dissipation
instrument 60 W 205 Btu/hr