Measure Protein A205
86 NanoDrop One User Guide Thermo Scientific
Related Topics
Beer-Lambert Equation
Protein A280 Calculations
Sample Pathlength
For micro-volume measurements, the software selects the optimal
pathlength (between 1.0 mm and 0.03 mm) based on sample
absorbance at the analysis wavelength.
For cuvette measurements, pathlength is determined by the cuvette
Pathlength setting in the software (see General Settings).
Displayed spectra and absorbance values are normalized to a 10 mm
pathlength equivalent.
Reported Values
Protein concentration. Reported in selected unit (mg/mL or μg/mL).
Calculations are based on Beer-Lambert equation using corrected
protein absorbance value.