46 AU-OPR-AureFloFT-EN,
Rev H
EMC Tables AureFlo System
Table 204 — Guidance and Manufacturers Declaration -
Emissions Equipment and Systems that are NOT Life-Supporting
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
The AureFlo® is intended for use in the electromagnetic environment specied below.
The customer or user of the AureFlo® should ensure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIRONMENT-GUIDANCE
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile RF communications equipment
should be no closer to any part of the AureFlo
®
including cables, than the recommended separation
distance calculated from the equations applicable to
the frequency of the transmitter.
Recommended separation distance:
d = [
3.5
]√P
V
1
d = [
3.5
]√P
E
1
d = [
7
]√P
E
1
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey,
a
should
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reection from structures,
objects and people.
a
Field strengths from xed transmitters, such as base stations fro radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the AureFlo
®
is
used exceeds the applicable RF compliance level above, the AureFlo
®
should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the AureFlo
®
.
b
Over the frequency rang 150 kHz to 80 MHz, eld strengths should be less than [V
1
] V/m.

Table of Contents