6–18
Bottom Density – Measurement Entry
Bottom density values may be obtained by using the gauge to record an
actual measurement. Follow the instructions from the Bottom Density –
Keypad Entry description on the previous page until reaching the Select
method of... display.
Press 2 for measurement entry. The display will be:
# of Readings to
average (1-20)?
x
Press ENTER
Input the number of readings to be taken and averaged. Press
START/ENTER for the display:
- Reading # 1 -
Time: x min.
Set rod to BS &
Press START
Prepare the site for measurement. Position the gauge and press
START/ENTER. The display will be:
- NOMOGRAPH -
Reading # 1
Depth: BS
Time:xx sec.
After counting down to zero, the display will be:
Reading # 1
WD:xxxxx PCF
Move gauge and
press START
If more than one reading is to be taken, move the gauge to the next location
and press START/ENTER. The gauge will repeat the above sequence
for each reading. When all the readings are completed, the display will be:
Average of x
Reading(s)-
xxxxx PCF
Press ENTER

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler 3440-L and is the answer not in the manual?

Troxler 3440-L Specifications

General IconGeneral
BrandTroxler
Model3440-L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals