108
1 EXP. HOLD FUNCTIONS
108
EXP. HOLD FUNCTIONS
Chapter
1
12. A: Sample hold B: Inflection point hold
An arbitrary point is held when the SECTION signal is ON. When the Hold Start Level is detected
after turning-ON of the SECTION signal, Inflection Point hold detection starts.
For the Hold Start Level, Detection Start Condition can be selected.
Also, operation can be
performed without using the Hold Start Level with the hold Detection Start Condition External Only.
The value is maintained until the T/H signal turns ON.
13. A: Sample hold B: Relative (Maximum / Minimum / Difference) hold
An arbitrary point is held when the SECTION signal is ON.
When the Hold Start Level is detected after
turning-ON of the SECTION signal, Relative Maximum / Minimum / Difference hold detection starts.
For the Hold Start Level, Detection Start Condition can be selected.
Also, the value is maintained until the T/H signal turns ON with the hold Detection Start
Condition External Only. The value is maintained until the T/H signal turns ON.
When the Detection Start Condition is External Only
Sensor Input Value
Hold Value
A
Hold Value B
H/E
SECTION
ON
T/H
Status Display
Status Display
Blink
----(A)
OFF
HOLDA
Blink
Status Display
Status Display
ON OFF
INFB
OFF Blink ON OFF
HOLDB
OFF
If there is no inflection point in the
changes of the sensor input value,
resulting in no hold, both the HOLD
and INF status indications will flash.
Also, the H/E signal will not turn ON.
B : In the case of Relative Minimum hold
When the Detection Start Condition is External Only
Sensor Input Value
Hold Value
A
Hold Value B
H/E
SECTION
T/H
BlinkON
OFF
Status Display
Status Display
OFF
ON OFF
Status Display
HOLDA
Blink
Status Display
----(A)
MAXB*
OFF
HOLDB
OFF
Blink ON
If there is no relative maximum/
minimum / difference in the changes
of the sensor input value, resulting in
no hold, both the HOLD and MIN*
status indications will flash.
Also, the H/E signal will not turn ON.
* MAX for Relative Maximum, and
DIF for Relative Difference.