9. Cleaning and Maintenance
Maintenance Tasks
Maintenance
period
Ensure that the power cable is not kinked, broken, exposed
and that it is free of all other defects.
Weekly
Conrm that activating the emergency stop button ceases
the unit’s operation. Pressing the emergency stop button
should stop all equipment movements and X-Ray emission.
Weekly
Ensure that all visible labels are intact and legible. Monthly
Check for possible wear or damage to the exposure switch
cable.
Monthly
Conrm that the audio message is audible throughout the
duration of the exposure.
Monthly
Pi_u272_en_20180514.indd 106 2018-05-14 오전 11:03:30

Table of Contents

Other manuals for Vatech PAX-I

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PAX-I and is the answer not in the manual?

Vatech PAX-I Specifications

General IconGeneral
Focal Spot Size0.5 mm
Image ReceptorDigital sensor
CompatibilityDICOM 3.0
Image DetectorCMOS
Power Supply220-240 VAC, 50/60 Hz
Sensor TypeCMOS
X-ray GeneratorHigh Frequency
Tube Voltage60-90 kVp
Tube Current4-12 mA
Power Requirements100-240V AC, 50/60Hz
UsageDental diagnostics