PaX-i (PCH-2500) User Manual
125
Appendices
2.3
X-Ray Scatter Dose
Test Equipment Information
Instrument Manufacturer Model S / N
Dose Meter Victoreen 660
101114 /
101377
PANO Mode
Test Condition
Mode Tested
PANO UHD
Distance from focal point [m]
1
Applied Tube Voltage Peak [kVp]
90
Applied Tube Current [mA]
10
Mode
Direction [°]
PANO UHD, 16.6 s [mR/hr]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
0 15.3 9.76 2.32
45 35.3 22.9 16.5
90 70.6 40.8 21.8
135 387 237 106
180 408 266 114
Mode
Direction [°]
PANO UHD, 16.6 s [mR/hr]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
225 166 63.9 38.8
270 29.5 13.4 7.74
315 19.6 10.1 6.33
Pi_u272_en_20180514.indd 125 2018-05-14 오전 11:03:33

Table of Contents

Other manuals for Vatech PAX-I

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PAX-I and is the answer not in the manual?

Vatech PAX-I Specifications

General IconGeneral
Focal Spot Size0.5 mm
Image ReceptorDigital sensor
CompatibilityDICOM 3.0
Image DetectorCMOS
Power Supply220-240 VAC, 50/60 Hz
Sensor TypeCMOS
X-ray GeneratorHigh Frequency
Tube Voltage60-90 kVp
Tube Current4-12 mA
Power Requirements100-240V AC, 50/60Hz