PaX-i (PCH-2500) User Manual
125
Appendices
2.3
X-Ray Scatter Dose
Test Equipment Information
Instrument Manufacturer Model S / N
Dose Meter Victoreen 660
101114 /
101377
PANO Mode
Test Condition
Mode Tested
PANO UHD
Distance from focal point [m]
1
Applied Tube Voltage Peak [kVp]
90
Applied Tube Current [mA]
10
Mode
Direction [°]
PANO UHD, 16.6 s [mR/hr]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
0 15.3 9.76 2.32
45 35.3 22.9 16.5
90 70.6 40.8 21.8
135 387 237 106
180 408 266 114
Mode
Direction [°]
PANO UHD, 16.6 s [mR/hr]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
225 166 63.9 38.8
270 29.5 13.4 7.74
315 19.6 10.1 6.33
Pi_u272_en_20180514.indd 125 2018-05-14 오전 11:03:33