PaX-i (PCH-2500) User Manual
126
Appendices
3.
Electromagnetic Compatibility
(EMC) Information
Guidance and manufacturer’s declaration - electromagnetic emissions.
The model PaX-i(PCH-2500) is intended for use in the electromagnetic
environment specied below. The customer or the user of the model PaX-
i(PCH-2500) should assure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
The model PaX-i(PCH-2500)
uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A (The model
PaX-i(PCH-2500)
in combination
with the shielded
location)
The model PaX-i(PCH-2500) must
be used only in a shielded location
with a minimum RF shielding
effectiveness and, for each cable
that exits the shielded location, a
minimum RF filter attenuation of
20 dB from 30 MHz to 230 MHz,
20 dB from 230 MHz to 1 GHz.
The model PaX-i(PCH-2500) ,
when installed in such a shielded
location, is suitable for use in
all establishments other than
domestic, and may be used in
domestic establishments and those
directly connected to the public
low-voltage power supply network
that supplies buildings used for
domestic purposes.
Harmonic
emissions IEC
61000-3-2
Not applicable
Voltage
uctuations
/ icker
emissions IEC
61000-3-3
Not applicable
NOTE) It is essential that the actual RF shielding effectiveness and filter
attenuation of the shielded location be veried to ensure that they meet or
exceed the specied minimum values.
Pi_u272_en_20180514.indd 126 2018-05-14 오전 11:03:33

Table of Contents

Other manuals for Vatech PAX-I

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PAX-I and is the answer not in the manual?

Vatech PAX-I Specifications

General IconGeneral
Focal Spot Size0.5 mm
Image ReceptorDigital sensor
CompatibilityDICOM 3.0
Image DetectorCMOS
Power Supply220-240 VAC, 50/60 Hz
Sensor TypeCMOS
X-ray GeneratorHigh Frequency
Tube Voltage60-90 kVp
Tube Current4-12 mA
Power Requirements100-240V AC, 50/60Hz
UsageDental diagnostics