PaX-i (PCH-2500) User Manual
122
Appendices
2.
X-Ray Dose Data
The X-Ray dose data is extracted from the X-Ray Dose Test Report for the
PaX-i.
X-Ray Dose Test Report for the PaX-i maintains dosemetric evaluation of
VATECH dental diagnostic system meets all requirements specied in the IEC
Collateral Standard. To limit unnecessary exposure to the patient, operator or
other staff, the PaX-i is designed to comply with IEC 60601-1-3 Part 1 General
Requirements for Safety.
Test Condition
Model (Brand Name)
PCH-2500 (PaX-i)
Sensor type
PANO: Xmaru 1501CF
CEPH: Xmaru 2301CF
Tube type / Inverter type
EXG9
2.1
DAP Table
Testing Equipment
Instrument Manufacturer Model S/N
Dose Meter Piranha 255 CB2-08050121
Mode Tested: PANO HD Normal, 13.5 s
mA
4 5 8
[mGy] [mGy
ㆍ
cm2] [mGy] [mGy
ㆍ
cm2] [mGy] [mGy
ㆍ
cm2]
60
5.01 33.5 7.79 52.0 10.26 68.4
70
7.12 47.5 10.41 69.5 13.72 91.6
80
9.06 60.5 13.26 88.5 17.48 116.7
90
11.12 74.2 16.28 108.7 21.45 143.1
kVp
Pi_u272_en_20180514.indd 122 2018-05-14 오전 11:03:32