EasyManuals Logo

Vatech PAX-I User Manual

Vatech PAX-I
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #124 background imageLoading...
Page #124 background image
PaX-i (PCH-2500) User Manual
124
Appendices
Mode
Direction [°]
PANO
(HD, Adult)
PANO
(HD, Child)
CEPH
(Scan type)
CEPH
(One-shot type)
[mR/hr]
120 23.8 15.5 25.1 30.4
130 17.4 15.0 20.6 26.6
140 17.3 16.2 17.6 30.2
150 18.5 17.0 22.1 33.7
160 21.2 17.5 23.5 67.8
170 23.0 16.7 25.5 36.8
180 6.1 4.7 5.2 19.9
190 16.4 10.8 16.9 26.9
200 20.7 18.2 23.7 30.6
210 50.7 13.3 15.6 29.3
220 74.8 12.9 11.2 26.8
230 70.6 11.3 10.1 21.5
240 60.1 11.7 11.8 16.0
250 35.1 17.5 12.9 12.8
260 19.2 11.9 11.3 14.2
270 12.9 10.9 10.6 11.3
315 10.3 8.0 9.0 10.4
Pi_u272_en_20180514.indd 124 2018-05-14 오전 11:03:32

Table of Contents

Other manuals for Vatech PAX-I

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PAX-I and is the answer not in the manual?

Vatech PAX-I Specifications

General IconGeneral
Focal Spot Size0.5 mm
Image ReceptorDigital sensor
CompatibilityDICOM 3.0
Image DetectorCMOS
Power Supply220-240 VAC, 50/60 Hz
Sensor TypeCMOS
X-ray GeneratorHigh Frequency
Tube Voltage60-90 kVp
Tube Current4-12 mA
Power Requirements100-240V AC, 50/60Hz
UsageDental diagnostics