6. Acquiring PANO Images
F. Ask the patient to:
Stand upright
Firmly grip the handle
Lean his/her chest lightly against the
equipment
Position his/her feet slightly forward and
open stance
G. Ensure that the patient’s shoulders
remain level and that his/her neck is
relaxed. The cervical spine should be
straight and upright.
H. Have the patient bite the bite block
along its grooves using his/her upper and
lower incisors (for edentulous patients, use
the edentulous chin support).
I. Have the patient
Close his/her lips around the bite block
Keep his/her tongue pressed against
his/her palate
Close his/her eyes
Ask the patient to remain still until
scanning is completed.
CAUTION
To acquire the best image possible, ask the patient not to:
- Breathe or swallow saliva during image acquisition
- Move during image acquisition
Pi_u272_en_20180514.indd 70 2018-05-14 오전 11:03:24

Table of Contents

Other manuals for Vatech PAX-I

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PAX-I and is the answer not in the manual?

Vatech PAX-I Specifications

General IconGeneral
Focal Spot Size0.5 mm
Image ReceptorDigital sensor
CompatibilityDICOM 3.0
Image DetectorCMOS
Power Supply220-240 VAC, 50/60 Hz
Sensor TypeCMOS
X-ray GeneratorHigh Frequency
Tube Voltage60-90 kVp
Tube Current4-12 mA
Power Requirements100-240V AC, 50/60Hz
UsageDental diagnostics