PaX-i (PCH-2500) User Manual
79
6. Acquiring PANO Images
WARNING
Be careful not to project the laser beams directly into the
patient’s eyes as this could severely damage the patient’s
vision.
A.
Mid-sagittal plane laser beam:
Position the Mid-sagittal plane laser
beam at the center of the patient’s face to
prevent magnifications of the left or right
side in the nal image.
B.
Frankfurt plane laser beam:
The
Frankfurt plane laser beam should go from
the top of the ear to the tip of the nose
when the patient’s head is tilted back 10 –
15 °.
You may adjust the Frankfurt plane laser
beam by manually adjusting the Frankfurt
plane laser beam lever.
C.
Canine laser beam:
Have the patient
smile to properly position the canine laser
beam on the center of the patient’s canine
tooth.
Completing Patient Positioning
A. After conrming patient positioning and beam alignment, adjust the temple
supports to fit snugly on both sides of the patient’s head using the temple
support wheel. The temple support wheel is located underneath the patient
support rest.
B. Click the
Ready
button on the PC or Touchscreen after the patient has been
properly positioned. At this point, X-Ray will not be emitted. The rotating unit
will move to position for imaging.
C. Proceed to section
6.4: Initiating X-Ray Exposure.
Canine laser beam
Mid-sagittal plane
laser beam
Adjusting lever for the
canine
Frankfurt plane laser
beam
Pi_u272_en_20180514.indd 79 2018-05-14 오전 11:03:26

Table of Contents

Other manuals for Vatech PAX-I

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PAX-I and is the answer not in the manual?

Vatech PAX-I Specifications

General IconGeneral
Focal Spot Size0.5 mm
Image ReceptorDigital sensor
CompatibilityDICOM 3.0
Image DetectorCMOS
Power Supply220-240 VAC, 50/60 Hz
Sensor TypeCMOS
X-ray GeneratorHigh Frequency
Tube Voltage60-90 kVp
Tube Current4-12 mA
Power Requirements100-240V AC, 50/60Hz
UsageDental diagnostics

Related product manuals