PaX-i (PCH-2500) User Manual
93
7. Acquiring CEPH Images
E. Ask the patient to stand upright facing
the sensor.
Make sure that the patient’s shoulders are
level and that his/her neck is relaxed.
F. Adjust the height of the unit to suit the
patient by pressing the column up/down
button or switch (optional).
After adjusting the column to the patient’s height, fit the
ear rods along the patient’s ear canals.
G. The ear rods should t properly along
the patient’s ear canals. The patient’s
Frankfurt Plane should be parallel to the
oor.
H. Ask the patient to swallow saliva and remain still with his/her mouth closed
until X-Ray exposure is complete.
I. Click the
Ready
button after the patient has been positioned. No X-Ray will
be emitted at this point. The CEPH sensor and the 2
nd
collimator will enter
capturing mode.
J. Proceed to section
7.4: Initiating X-Ray Exposure.
Sensor
Frankfurt plane