Operation 4–10
Select O/C Trim using the Navigation Keys and then press Enter. The following screen will be
displayed:
Figure 4-8 Open Circuit Trim Page
1) Open-circuit the Kelvin clips (see Fig 4-5), and press Enter. If required, use the Navigation
Keys to select Cancel and then press Enter.
2) The instrument will display the following message while the open-circuit trim routine is
being performed:
Figure 4-9 Open Circuit Trim in progress
3) When the O/C trim has been completed, the instrument will return to its original screen.
4) Select O/C Trim using the Navigation Keys and then press Enter. The following screen
will be displayed: