4300 LCR Meter Specification
7–4
7.5 Modes Of Operation
7.5.1 Measurement
Single and repetitive measurements of any two parameters displayed as Function 1 and
Function 2.
Analogue scale with configurable Hi/Lo limits giving PASS/FAIL indication.
7.5.2 Binning (Option)
B1 version: All signals are non-isolated
B2 version: All signals are isolated using opto-couplers
Test result outputs and control outputs are identical for B1 and B2 versions
7.5.2.1 CONTROL INPUT:
B1 version:
TRIG
B2 version: Trigger input +ve pin
Trigger input –ve pin
7.5.2.2 TEST RESULT OUTPUT:
5 BIN’s (numbered 0 to 4) for valid measurements
BIN 9 used for O/R (Out of Range) measurements
Pass/Fail output
7.5.2.3 CONTROL OUTPUT:
BUSY
(Binning Data Available)
7.5.2.4 ELECTRICAL RATINGS
B1 version (non-isolated):
Parameter Specification
Output High: >4.0V
Output Low: <1.0V
Input High: >3.5V
Input Low: <1.5V
Drive capability (low) 10mA sink
Drive capability (high) 30µA source