EasyManua.ls Logo

Wayne 4310 - Binning - Count; Figure 4-52 Count Screen Example

Wayne 4310
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation 4–36
4.4.26 Binning - Count
This section is used to display the results based on the binning performed using the Binning
Sort section. It is accessed by selecting Count from the Binning screen of the More pull down
menu (see 4.4.23).
The following is an example of a typical screen:
Figure 4-52 Count screen example
The total number of measurements made is shown as Total (5 in this example).
The reject quantity is calculated as follows:
One Term: Reject = Total – BIN 0
One Term Dual Limits: Reject = Total – (BIN 0 + BIN 3 + BIN 4)
Two Term: Reject = Total – BIN 0
The total number of DUT’s allocated to each individual bin is also displayed, together with the
Binning Type and measurement conditions.
The Del Last soft key (Delete Last) can be used to reduce the total count by 1. This can be
useful if a known bad measurement (due to a poor connection for example) is made, and it
needs to be removed from the counts. This can only be done once after each trigger.
The Del All soft key (Delete All) can be used to reset all binning counts to 0. A warning
message will ask the user to confirm this reset before it is performed.