Operation 4–15
The following features can be selected from the Set Pull Down Menu:
Function 1
This allows any one of the following 11 AC measurement parameters to be
selected.
C Capacitance
L Inductance
X Reactance
B Susceptance
Z Impedance
Y Admittance
Q Quality Factor
D Dissipation Factor (tan delta)
R AC Resistance
G Conductance
A Phase Angle
It also allows Rdc (DC resistance) to be selected when this option is fitted.
Function 1 can also be selected using the Soft Key labeled Func 1.
Function 2
This allows any of the 11 AC measurement parameters offered in
Function 1 to be selected (but not Rdc).
Function 2 can also be selected using the Soft Key labeled Func 2.
Equ. Cct.
This allows the Equivalent Circuit
used to calculate the AC measurements
to be toggled between Series and Parallel
Frequency
This allows the AC measurement frequency to be set between 20Hz and
100kHz for 4310
200kHz for 4320
500kHz for 4350
1MHz for 43100
Frequency can also be selected using the Soft Key labeled Freq.
Drive Level
This allows the AC Drive Level to be selected between 10mVrms and
2Vrms.
Drive Level can also be selected using the Soft Key labeled Level.