REMOTE CONTROL (GPIB & RS-232) 5–14
MEASUREMENT MODE
:MEAS:TRIGger
Trigger a measurement using the current settings.
Parameters:
None.
Response:
For AC measurements the response will be the first and second measurements separated by
a comma.
Example: 68.860E-9 , 13.0E+6
For Rdc measurements the response will be a single measurement result.
Example: 6.2295E+3
:MEAS:FREQuency <real>
Set frequency of AC measurement.
Parameters:
The required frequency in Hertz. The
unit suffix ‘Hz’ is optional.
Example: MEAS:FREQ 1000
sets the test frequency to 1kHz.
Response:
None.
:MEAS:FREQuency?
Returns the current AC test frequency.
Parameters:
None.
Response:
Returns the current test frequency in
engineering format.
Example: +.10000000E+04
for a test frequency of 1kHz.
:MEAS:LEVel <real>
Set drive level for currently selected test.
Parameters:
For AC tests supply the required drive
level in either Volts or Amps.
Example: MEAS:LEV 1.2
will select drive levels of 1.2V.
Response:
None.
:MEAS:LEVel?
Drive level query.
Parameters:
None.
Response:
Returns the current test level in
engineering format.
Example: +.20000000E-01
for a test level of 20mV.