REMOTE CONTROL (GPIB & RS-232) 5–15
MEASUREMENT MODE
:MEAS:SPEED <disc>
Select the required measurement speed.
Parameters:
MAX Maximum speed.
FAST Fast speed.
MED Medium speed.
SLOW Slow speed.
Example: :MEAS:SPEED SLOW
will select slow speed for
measurements.
Response:
None.
:MEAS:SPEED?
Returns the current test speed.
Parameters:
None.
Response:
Returns the test speed as an integer
according to the table:
0 Maximum
1 Fast
2 Medium
3 Slow
Example: 1
indicates that Fast measurements are
selected.
:MEAS:RANGE <disc>
Select the required measurement range
condition for AC and RDC tests.
Parameters:
The following parameters are valid:
AUTO Auto-ranging.
1 to 8 Range 1 to 8
Example: MEAS:RANGE 1
MEAS:RANGE AUTO
will select range 1 and auto-ranging
respectively.
Response:
None.
:MEAS:RANGE?
Returns the current measurement range.
Parameters:
None.
Response:
Returns the measurement range as an
integer according to this table:
0 Auto-ranging.
1-8 Current measurement range.
Example: 0
indicates that auto ranging is selected.