Operation 4–33
The Low Limit soft key is used to enter LL1 and LL2 limits. These are set in the same way as
the High Limits described above.
The Nom(inal) soft key is only displayed when the Abs/Perc soft key is toggled to Perc. It is
used to enter the nominal value which the percentage limits are calculated from. After pressing
the Nom soft key, the nominal value can be entered in the same way as High Limits and Low
Limits. It is also possible to save the last measurement as the Nominal Value. This can be done
by pressing the Save Res(ult) soft key which appears after pressing the Nom soft key. When
absolute limits are selected, the word “Nom” is still shown on the screen, but is not used.
Figure 4-47 One Term Dual Limits Save Result screen
When the limits have been set, measurements and binning can be started from either the Sort or
Count screens.
4.4.24.3 Binning – Setting Two Term Limits
One of the following two screens will be displayed when the Two Term Binning Type has
been selected, depending on whether the last limits used were Absolute or Percentage.
Figure 4-48 Two Term Absolute limits screen