4-9
IM 760301-01E
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Measurement Conditions
• Power Range
The measurement ranges (power ranges) of active power (P), apparent power (S),
and reactive power (Q) are as follows:
Wiring System Power Range
1P2W (single-phase, two-wire) voltage range × current range
1P3W (single-phase, three-wire) voltage range × current range × 2
3P3W (three-phase, three-wire) (when the voltage and current ranges on the corresponding
elements are set to the same range)
3P3W (3V3A, three-voltage, three-current)
3P4W (three-phase, four-wire) voltage range × current range × 3
(when the voltage and current ranges on the corresponding
elements are set to the same range)
• When the result of the equation
voltage range
×
current range
exceeds 1000 W
(VA or var), the displayed unit changes to kW (kVA or kvar).
• The display resolution is 600000.
Note
Because the voltage and current ranges switch independently according to range increase
and decrease conditions in auto range mode, different power ranges may be set for the same
power value.
4.3 Setting the Measurement Range for Direct Input

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA wt3000 and is the answer not in the manual?

YOKOGAWA wt3000 Specifications

General IconGeneral
BrandYOKOGAWA
Modelwt3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals