EasyManuals Logo
Home>YOKOGAWA>Measuring Instruments>wt3000

YOKOGAWA wt3000 User Manual

YOKOGAWA wt3000
384 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #170 background imageLoading...
Page #170 background image
5-34 IM 760301-01E
• Limitation on Modifying the Settings during Integration
When the integration is in progress, settings of some functions cannot be changed as
shown below.
Function
IntegratingIntegration Reset
ON
OFF
No
No
No
No
No
No
Yes
Yes
No
Yes
No
(Displayable)
No
(Displayable)
No
Yes
No
Yes
No
No
No
OFF
OFF
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
Yes
Yes
Yes
Yes
Yes
Wiring system
Measurement range
Scaling
Filter
Averaging
Synchronization source
Hold
Single measurement
Data update rate
Display mode
Integration mode
Integration timer
Integration start
Integration stop
Integration reset
Store (excludes integration
synchronization mode)
Printer
Zero-level crossing
Null
(START indicator)
(STOP indicator)
Yes: Settings can be modified.
No: Settings cannot be modified.
Starting integration in auto range mode switches the measurement range to fixed range mode.
Integration Suspended
OFF
ON
No
No
No
No
No
No
Yes
Yes
No
Yes
No
(Displayable)
No
(Displayable)
Yes
No
Yes
Yes
Yes
No
No
Integration Operation Condition
• Limitation on the Waveform Display Function during Integration
The trigger function (see section 6.3) for waveform display does not work while
integration is in progress or is suspended. Therefore, the signal level of the waveform
display at the left end of the screen may not be stable.
5.11 Integration