10-3
IM 760301-01E
1
2
3
4
5
6
7
8
9
10
10
11
12
App
Index
Other Functions
Measurement Functions Affected by the NULL Function
The NULL value is subtracted from the sampled data. Hence, all measurement
functions are affected by the NULL values.
Limitations on the NULL Function by Measurement Modes
The NULL function cannot be used in the following measurement modes.
Wide bandwidth harmonic measurement mode
IEC harmonic measurement mode
Voltage fluctuation and flicker measurement mode
Cycle-by-cycle measurement mode
10.2 NULL Function