App-4
IM WT5000-02EN
Code Message Chapter or Section
849 Attempted made to start storing in real time counting storing mode when the end time had
already passed.
7.1
850 Cannot be set or executed at current store state.
To set or execute, reset store.
7.4
852 Stored file is illegal.
Initialize memory before storing.
7.4
854 Waveform display data not found. —
855 Data destination memory is full.
Saving has been stopped.
—
856 An error has occurred while storing.
Storing has been stopped.
—
857 Cannot be set while Master/Slave Synchronized
Measurement is set to Slave.
2.14
858 Store process is in progress now.
Execute or set setting again.
—
859 Cannot convert selected file.
Select a file with an extension of WTS or HDS.
7.3
862 Numeric data not found. —
863 Cannot be set or executed when different types of elements are installed. —
864 This wiring cannot be set as the first selected element. 2.1
865 Cannot be set while integration is running.
Stop or reset Integration.
4.2
874 Sync source, PLL source or trigger source cannot be set to Ext Clk, while Master/Slave
Synchronization Measurement is set to Slave.
2.14
875 Master/Slave Synchronization Measurement cannot be set to Slave, while sync source,
PLL source or trigger source is set to Ext Clk.
2.14
876 Can not calculate from present point value. 9.1
877 Can not set 0 to count. 7.1
878 Can not set 0. Set to a value other than 0. —
880 Cannot be set or executed while initialization.
Wait until status changes to "Ready".
—
881 Cannot be set or executed while measurement is in progress.
To set or execute, "Stop" measurement.
—
882 Stopped measurement. Detection error of measuring interval signal. Check External Sync
(MEAS START) input.
4.3*
885 Cannot be set or executed in Current Measurement Mode.
Set or execute in Normal Measurement Mode.
1.1
889 Setting and execution is not. available when auto-ranging is set to ON. —
893 Invalid cutoff frequency setup. Set the cutoff frequencies so that the low pass filter's (LPF)
cutoff frequency is higher than the high pass filter (HPF).
2.7
894 Cannot be executed/set in current condition.
For all current inputs with in the same wiring, “CT Type” or “Input Resistance” must have
the same setup.
2.1
* Getting Started Guide, IM WT5000-03EN
Appendix 1 Messages and Corrective Actions