SonTek/YSI
34
ADVField Operation Manual (September 1, 2001)
SyncMode or SM [DISABLE|START|SAMPLE]
Default parameter:
DISABLE
Without parameter: returns its current setting.
When used with a valid parameter, sets the external synchronization mode. The ADVField
processing module includes a sync input line (SyncIn) to allow external hardware control of
ADV data collection.
DISABLE
indicates that data collection is completely under internal software control.
START
indicates that data collection starts with the rising edge of the first input sync
pulse and continues at the software-programmed sampling rate.
SAMPLE
indicates that data collection starts with the rising edge of the first input sync
pulse and continues at a rate determined by the time between the rising edge of succes-
sive input sync pulses.
See §5.5 for information about external synchronization.