SamplesPerBurst or SPB [d] [d] [d]
Default parameters:
1200 0 0
Parameter range:
1
to
32000
Without parameter: returns its current settings for each burst type.
When used with from one to three valid parameters, sets the number of samples per burst for
each burst type. See §1.3 for a description of burst sampling.
For autonomous deployments, the ADVField supports up to three samples per burst values
for three burst types. Samples per burst are not used during real-time data collection.
To select multiple-burst sampling, enter a separate samples-per-burst for each burst type de-
sired. Example:
SPB 24 600 7500
sets the samples per burst for burst type #1 to 24, for
burst type #2 to 600, for burst type #3 to 7500. If only one or two input arguments are speci-
fied, the values for the other burst types are set to 0, disabling that burst type.
RecordAmpCorr [YES|NO] [YES|NO] [YES|NO]
Default parameters:
YES YES YES
Without parameter: returns its current settings for each burst type.
When used with from one to three valid parameters, tells the ADV whether to record ampli-
tude and correlation data. See §1.3 for a description of burst sampling.
YES
tells the ADV to record amplitude and correlation data for each receiver with each
sample for the selected burst types.
YES
increases the size of each sample by 4 bytes.
NO
tells the ADV to record only the mean amplitude (from all 3 receivers) and the mean
correlation (from all 3 receivers) with each sample for the selected burst types.
This parameter is used only in deployment mode; it has no affect on real-time data collection.
If only one or two input arguments are specified, it does not change the settings for the non-
specified burst types.
Appendix 2 has a description of the data recorded based on the
Record
command settings.
RecordCompass [YES|NO] [YES|NO] [YES|NO]
Default parameters:
No
compass/tilt sensor installed:
NO NO NO
With compass/tile sensor installed:
YES YES YES
Without parameter: returns its current settings for each burst type.
When used with from one to three valid parameters, tells the ADV whether to record com-
pass/tilt sensor data. See §1.3 for a description of burst sampling.
YES
tells the ADV to record compass/tilt sensor data (heading, pitch, roll) for each veloc-
ity sample for the selected burst types.
YES
increases the size of each sample by 6 bytes.
NO
tells the ADV not to record compass/tilt sensor data for the selected burst types.
Regardless of this setting, the ADV will always record velocity data in either XYZ or
ENU coordinates based on the setting of the
CoordSystem
setup parameter (see §3.6).
This parameter is used only in deployment mode; it has no affect on real-time data collection.
If only one or two input arguments are specified, it does not change the settings for the non-
specified burst types.
Appendix 2 has a description of the data recorded based on the