EasyManuals Logo

YSI SonTek ADVField User Manual

YSI SonTek ADVField
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001) 39
RecordExtPress [YES|NO] [YES|NO] [YES|NO]
Default parameters:
NO NO NO
Without parameter: returns its current settings for each burst type.
When used with from one to three valid parameters, tells the ADV whether to record data
from an optional, external pressure sensor. See §1.3 for a description of burst sampling.
YES
tells the ADV to record external pressure sensor data for each velocity sample for
the selected burst types.
YES
increases the size of each sample by 3 bytes.
NO
tells the ADV not to record external pressure sensor data for the selected burst types.
This parameter is used only in deployment mode; it has no affect on real-time data collection.
If only one or two input arguments are specified, it does not change the settings for the non-
specified burst types.
Appendix 2 has a description of the data recorded based on the
Record
command settings.
RecordCTD [YES|NO] [YES|NO] [YES|NO]
Default parameters:
NO NO NO
Without parameter: returns its current settings for each burst type.
When used with from one to three valid parameters, tells the ADV whether to record data
from an optional CTD sensor. See §1.3 for a description of burst sampling.
YES
tells the ADV to record optional CTD data for each velocity sample for the selected
burst types.
YES
increases the size of each sample by x bytes, where x is dependent on the
type of CTD sensor being used.
NO
tells the ADV not to record CTD data for the selected burst types.
This parameter is used only in deployment mode; it has no affect on real-time data collection.
If only one or two input arguments are specified, it does not change the settings for the non-
specified burst types.
Appendix 2 has a description of the data recorded based on the
Record
command settings.
RecordLisst [YES|NO] [YES|NO] [YES|NO]
Default parameters:
NO NO NO
Without parameter: returns its current settings for each burst type.
When used with from one to three valid parameters, tells the ADV whether to record data
from an optional LISST-100 sensor. See §1.3 for a description of burst sampling.
YES
tells the ADV to record optional LISST data for each velocity sample for the se-
lected burst types.
YES
increases the size of each sample by 80 bytes.
NO
tells the ADV not to record LISST data for the selected burst types.
This parameter is used only in deployment mode; it has no affect on real-time data collection.
If only one or two input arguments are specified, it does not change the settings for the non-
specified burst types.
Appendix 2 has a description of the data recorded based on the
Record
command settings.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YSI SonTek ADVField and is the answer not in the manual?

YSI SonTek ADVField Specifications

General IconGeneral
BrandYSI
ModelSonTek ADVField
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals