EasyManua.ls Logo

YSI SonTek ADVField

YSI SonTek ADVField
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001) 71
standoffs; board orientation does not matter. Above the sensor board are four standoffs (#4-40 x
1/2”). Last is the compass, secured by four #4-40 x 3/8” bolts, with fiber washers on each side.
Align the compass as shown in Figure 21. For side-looking operation, it is not possible to align
the compass X-axis with the probe X-axis. Instead, the probe X-axis is aligned with the compass
negative Y-axis. Assuming the orientation switch is set for side-looking operation (§4.2), this
alignment is accounted for internally.
In side-looking operation, you must
ensure the probe’s X-axis (stamped
on the end cap and defined by re-
ceiver arm #1) is in the horizontal
plane (within the ±50° limits of the
tilt sensor). When looking at the rear
end cap, the pressure sensor should
be in the position shown in Figure 22.
If the system does not have a pressure
sensor, make sure the X-axis points
to the right of the underwater connec-
tor. Figure 22 shows the orientation.
If the system is not oriented correctly,
the compass will be upside down and
will give invalid data.
Probe X
X
Z
Compass Axis
Compass/Tilt
Sensor
3/8" Hex Standoffs
Pressure
/
Temp
Sensor Board
X
Y
Compass Axis
Figure 21 - Compass/Tilt Sensor, Side-Looking Mounting
Receiver #1
Vertically Up
Pressure
Underwater
Mbl
Connector
Figure 22 - Probe Mounting, Side-Looking Compass

Table of Contents

Related product manuals