HFA3 Instructions for Use 2660021166131 Rev. A 2018-11
Go to Contents Networking
7-7
6. Enable MWL and enter these values in the next screen:
A. Enter EmrSCP under AE title.
B. Enter localhost under Host name.
C. Enter 2112 under Port.
D. Select OK.
7. Ensure that the EMR is integrated into the shared/FTP export and worklist folders before
selecting Test connection on the instrument. A check mark will appear if the connection is
successful.
8. Do not change the default values under Specific settings.
Enabled services display a check mark to the left.
Troubleshooting the Connection
If the connection is unsuccessful, take the following actions:
Check with your network administrator to verify that the instrument is connected to the
network.
•Retry.
Make sure the appropriate servers are running.
The storage provider hard drive(s) may be full. Add more hard drives or make more space
available.
Refer to the DICOM/EMR system manual for a possible resolution of the problem.
Call Customer Service.

Table of Contents

Other manuals for Zeiss Humphrey Field Analyzer 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Zeiss Humphrey Field Analyzer 3 and is the answer not in the manual?

Zeiss Humphrey Field Analyzer 3 Specifications

General IconGeneral
BrandZeiss
ModelHumphrey Field Analyzer 3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals